INSTRUCTION MANUALDetcon Model TP-524D-HRTPGM 1PGM 2MODEL TP-524DHOUSTON, TEXASFLT 1 2 CALMicroSafe H2S Gas SensorALM ALMTMTP-524D-HRT Hydrogen Sulfid
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 6 of 48There are some gases typically found in industrial environments that can cause a
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 7 of 488-32 tappedground point4.6"34NPT Ports6.125"5.5"6.45"14"
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 8 of 48DrainConduit"T"EYS Seal FittingPGM 1PGM 2MODEL TP-524DHOUSTON, TEXASFLT
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 9 of 48NOTE 3:The supply of power should be from an isolated source with over-current pr
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 10 of 48c) Remove the desiccant cap about 10 minutes after applying power to the sensor
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 11 of 483. Operation3.1 Programming Magnet Operating InstructionsThe Operator Interface
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 12 of 483.2 Operator InterfaceThe operating interface is menu-driven via the two magneti
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 13 of 48Software FlowchartdecPGM1/2 (M)mA Output XX.XXSensor Temp XX CVoltage XX.X VDCRe
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 14 of 483.3 Normal OperationIn normal operation, the display continuously shows the curr
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 15 of 48concentration is 10ppm for the 0-20ppm range and 25ppm for the 0-50ppm and 0-100
Model TP-524D-HRTModel TP-524D-HRT iiThis page left intentionally blankShipping Address: 4055 Technology Forest Blvd, Suite 100, The Woodlands Texas 7
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 16 of 48NOTE 2: If the sensor fails the stability criteria, a“Stability Fault”will be de
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 17 of 48View Sensor Status displays all current configuration and operational parameters
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 18 of 48Alarm 1 LevelThe menu item appears as: “Alarm 1 Level XX”Alarm 1 AscendingThe me
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 19 of 48Set Range is used to change full-scale ranges. This is selectable between 0-20,
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 20 of 483.5.5 Signal Output CheckSignal Output Check provides a simulated 4-20mA output.
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 21 of 48 Heater Power: Heater Power settings are lost and “Set Heater Power” (3.5.4) mu
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 22 of 48indicates the relay is latching. Swipe the magnet momentarily over PGM2 or PGM1
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 23 of 48Range Fault – AutoSpanIf the sensor fails the minimum signal change criteria dur
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 24 of 48Input Voltage FaultIf the detector is currently receiving an input voltage that
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 25 of 484. HRT BridgeThe HRT Bridge interface is an integral part of the Transmitter Mod
Model TP-524D-HRTModel TP-524D-HRT iiiTable of Contents1. Introduction ...
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 26 of 48The HRT Bridge communicates with the TP-524D-HRT sensor through Modbus™. The HRT
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 27 of 48Figure 12 Primary VariablesPrimary Variables Concentration – the value of the g
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 28 of 484.4.1.2 IdentificationThe Identification screen contains 4 sections that provide
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 29 of 484.4.2 Diagnostics MenuThe Diagnostics Menu contains two screens; 1) Device Statu
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 30 of 48 Loop current fault – The sensor has detected a condition where the 4-20mA outp
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 31 of 48Figure 15 FP Configuration SetupThe Range Set up will display the Model Type of
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 32 of 48Figure 16 FP Calibration Screen4.4.3.3 Sensor Calibration using the HRT BridgeCa
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 33 of 48fall between 5% and 100% of range. However, any alternate span gas concentration
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 34 of 48“AutoSpan Complete”“Sensor Life XXX%”“Remove Span Gas”8. When the sensor passes
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 35 of 48c) Connect the positive lead of a DVM set to measure milliamps to the other end
Model TP-524D-HRTModel TP-524D-HRT ivTable of FiguresFigure 1 Sensor Assembly Front View ...
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 36 of 48Figure 18 HART Setup Tag – Text that is associated with the field device instal
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 37 of 485. Service and MaintenanceCalibration FrequencyIn most applications, monthly to
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 38 of 48Replacement of Transmitter Modulea) Remove the junction box cover and remove the
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 39 of 486. Troubleshooting GuideRefer to the list of Failsafe Diagnostic features listed
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 40 of 48Stability FaultProbable Causes: Failed Sensor, empty or close to empty Cal Gas C
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 41 of 48Restore Factory Defaults - This will clear the processor’s memory and may correc
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 42 of 487. Customer Support and Service PolicyDetcon HeadquartersShipping Address: 4055
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 43 of 488. TP-524D-HRT Sensor WarrantyH2S Sensor WarrantyDetcon Inc. warrants, under nor
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 44 of 489. Appendix9.1 SpecificationsSensor Type: Continuous diffusion/adsorption typeCH
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 45 of 48Storage Temperature: -35°C to +55°C (-31°F to +131°F)Operating Humidity: 5-100%
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 1 of 481. Introduction1.1 DescriptionDetcon Model TP-524D-HRT hydrogen sulfide sensors a
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 46 of 489.2 Spare Parts, Sensor Accessories, Calibration EquipmentPart Number Spare Part
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 47 of 489.3 Revision LogRevision Date Changes made Approval0.0 12/01/10 Initial Release
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 48 of 48Shipping Address: 4055 Technology Forest Blvd, Suite 100, The Woodlands Texas 77
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 2 of 48Principle of OperationMethod of detection is by diffusion/adsorption. Air and H2S
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 3 of 48Field Replaceable SensorThe Detcon solid-state H2S gas sensor is a field proven,
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 4 of 482. Installation2.1 Operational Guidelines for Safe Use1. Install sensor only in a
Model TP-524D-HRTTP-524D-HRT Instruction Manual Rev. 0.3 Page 5 of 48Normal ventilation or prevailing wind conditions can dictate efficient location o