and 24VDC source is shown in the Table below. The maximum wire size for termination in the Junction Box
is 14 AWG.
Table 1 Wire Gauge vs. Distance
AWG Wire Dia. Meters Feet
Protection
Wiring table is based on stranded tinned copper wire and is designed to serve as a
reference only.
Shielded cable is required for installations where cable trays or conduit runs include
high voltage lines or other possible sources of induced interference. Separate conduit runs are
highly recommended in these cases.
ld be from an isolated source with over
as stipulated in Table 1 Wire Gauge vs. Distance.
A 250 ohm load resistor is need
20 mA line when it is not being used.
Terminal Connections
CAUTION: Do not apply System power to the sensor until all wiring is properly terminated. Refer
to Section 2.8 Initial Start Up
Figure 11 Sensor Connector PCB
a) Remove the junction box cover and unplug the Transmitter Module. Identify the terminal blocks for
customer wire connections.
Customer
Wiring
White
Blue
Yellow
Black
Wiring to
LEL Sensor
Power (+ -)
4-20 mA
RS-485 (A,B)
Fault
Alarm 2
Alarm 1