Detcon FP-624D-HT Manual de usuario Pagina 45

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 48
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 44
Model FP-624D-HT
FP-624D-HT Instruction Manual Rev. 1.1 Page 41 of 44
Status Indicators: 4-digit LED Display with gas concentration, full-script menu
Prompts for AutoSpan, AutoZero, Set-up Options, and Fault Reporting
Faults Monitored: Loop Fault, Input Voltage Fault, Zero Fault, Sensor Fault,
Processor Fault, Memory Fault, Calibration Fault(s)
4 LEDs for Alarm 1, Alarm 2, Fault, and Calibration
Cable Requirements: Power/Analog: 3-wire shielded cable
Maximum distance is 13,300 feet with 14 AWG
Serial Output: 2-wire twisted-pair shielded cable specified for RS-485 use
Maximum distance is 4,000 feet to last sensor
Mechanical Specifications
Length: 10.85 inches (275mm), including Splashguard
Width: 6.1 inches (155mm)
Weight: 2.7lbs (3.4Kg)
Mechanical Connection: ¾” Male NPT threaded connection
Electrical Connection: Eleven 14 gauge (maximum) wire terminal landings